At a distance of two feet (0.61 metre) from the light source, the light output on the subject is 53000 Lux.
The colour temperature rendering of 3200 degrees K provides intense "daylight" illumination, with true tissue colour.
Three low-voltage 50 Watt halogen lamps are strategically positioned in order virtually to eliminate shadows, when a person or object is placed between the light source and the patient.
The Outpatient luminaire was designed and engineered with safety in mind. Continuous light is assured with the parallel wiring of the three lamps. If one lamp burns out the two remaining will still provide light. Additionally, the low-voltage operation minimizes electrical hazards during procedures.
The housing and light beam remain very cool, since the dichroic reflector filters out more infra-red rays than a standard reflector.
Each lamp has an average life of 2000 hours. | |||
|
COMMON FEATURES FOR BOTH COOLSPOT & OUTPATIENT LUMINAIRES
Both luminaires and fixtures operate on 240 Volts, 50Hz supply.
Both luminaires and fixtures are tested to BS 5724, SEMKO, NEMKO, DEMKO, and all major European Safety standards.
White | |
LUX MEASUREMENT AT SPECIFIED DISTANCES
COOLSPOT AND OUTPATIENT LUMINAIRES
Distance from subject Inches 24
30
39.5
Metres 0.61
0.76
1
Lux Level CoolSpot 47500
31500
13400
Outpatient 53000
51000
28000
The most typical usages distance from the subject is two feet (0.61 metre) IMPORTANT
Wall units must be securely mounted on walls which offer adequate support.
Ceilings must be able to support the weight of single and double ceiling mounts.
Further detailed mounting instructions are available for all product variations shown. It is recommended that these are studied at the specification stage.
Advice is available from the Technical Department at Luxo UK Ltd. or Harrison Veterinary Equipment.
![]()
![]()